NASA-HDBK-8739.19-4, NASA HANDBOOK: ESTIMATION AND EVALUATION OF MEASUREMENT DECISION RISK: NASA MEASUREMENT QUALITY ASSURANCE HANDBOOK – ANNEX 4 (13 JUL 2010)
NASA-HDBK-8739.19-4, NASA HANDBOOK: ESTIMATION AND EVALUATION OF MEASUREMENT DECISION RISK: NASA MEASUREMENT QUALITY ASSURANCE HANDBOOK – ANNEX 4 (13 JUL 2010)., This document is Annex 4 of NASA-HDBK-8730.19 [40]. It provides guidelines for computing
metrics by which conformance testing for test and calibration processes can be evaluated. Chief
among these metrics are measurement decision risks. These include the risk of incorrectly
accepting a nonconforming equipment attribute and the risk of incorrectly rejecting a conforming
attribute. The former is called False Accept Risk and the latter is termed False Reject Risk. A
principal objective of this Annex is to accumulate and focus the body of knowledge of
measurement decision risk to provide uniform guidance in the application of this knowledge to
estimate and manage such risk, particularly as it relates to NASA tests and calibrations.
In addition to providing guidelines for estimating and managing measurement decision risks, this
Annex addresses the evaluation of these risks in terms of their impact on both operating costs
and on the cost of the undesirable consequences of incorrect test decisions. Further, the
guidelines provided herein can be applied to equipment and system design issues involving
decisions regarding accuracy requirements and available support capability.
The principles and methods presented in this Annex are applicable to any conformance test that
relies on measured values to provide information on which to base acceptance or rejection
decisions.